Medical vacuum in hospitals — what it is for, how the vacuum source works and what the standard requires

2026-11-24 · Małgorzata Dopierała, Damian Czyczyro, Przemysław Kostera

In brief. Medical vacuum is negative pressure generated centrally by a vacuum source and carried through pipelines to terminal units, where it is used to suction fluids, secretions and gases from the treatment field. The vacuum supply system has at least a primary, a secondary and a reserve source, bacterial filters upstream of the pumps, a reserve in another room or fire compartment, and an exhaust discharged outside the building.

Medical vacuum works the opposite way to the other gases in a hospital. Nothing flows through it to the patient: the pipeline draws in air, and fluids suctioned at the bedside or in the operating theatre are held back by the jar and separator at the terminal unit. This determines how its source is designed and what its operation requires.

What is medical vacuum and what is it used for in a hospital?

Medical vacuum is negative pressure generated centrally by a vacuum source and maintained in the vacuum pipeline system. It is used to suction fluids, secretions and gases from the treatment field, at intensive care stations, in operating theatres and in other clinical areas.

The task of the vacuum source is to generate and maintain the required negative pressure, not to deliver gas to the patient. Biological material that gets past the terminal unit enters the pipeline. The vacuum source itself is part of the supply system of the medical gas installation, which is why GMWP says it must be treated as an element of that installation rather than as an ordinary technical pump room.

What does a medical vacuum installation consist of?

The vacuum installation is part of the pipeline system for medical gases and vacuum. It consists of the vacuum source, pipelines, zone control units, the monitoring and alarm system, and terminal units. Requirements for the source and pipelines are set by PN-EN ISO 7396-1, and for the terminal units by PN-EN ISO 9170-1.

The anaesthetic gas scavenging system (AGSS) is a separate system, covered by PN-EN ISO 7396-2. It has its own flow-generating equipment and is not connected to medical vacuum or to gas installations.

How is a medical vacuum source built?

A vacuum source is often a complete unit on a common supporting frame. The guide “Medical gases in practice. Volume 1” (GMWP), prepared by INMED S.A., describes an example unit with three pumps.

ElementTask
vacuum pumpsgenerate negative pressure; they may be e.g. side-channel, vane, claw, screw, lobe, liquid-ring or radial pumps
bacterial filtersprotect the pumps against biological and mechanical contamination drawn in from the installation
buffer tankstabilises the system and compensates short-term fluctuations in demand
vacuum sensor and vacuum gaugesend the vacuum level to the controls and alarms and show it locally
control cabinet with panelprotects the system electrically and shows the vacuum level, pump status, alarms and counters
service valvesallow a filter or pump to be replaced without shutting down the whole source
exhaust connectiondischarges air from the pumps outside the building

The buffer tank does not replace pump capacity or the reserve. If the system has too little capacity, the tank alone will not maintain vacuum during long or intensive use.

Why does a vacuum source need a reserve in another room?

As with oxygen and medical air, according to GMWP the vacuum supply system has at least a primary, a secondary and a reserve source. The primary source runs day to day, the secondary takes over when the primary is unavailable, and the reserve maintains negative pressure when both are unavailable.

Separating locations is especially important for vacuum. According to GMWP, at least one source or reserve section should be located in another room or another fire compartment, so that a fire, flooding, power failure, smoke, contamination or lack of access to the main pump room does not cut the hospital off from vacuum. If the reserve is started manually, access to it must also be ensured in an emergency.

How is the capacity of a vacuum source selected?

The source is sized to the hospital’s design flow. What counts is the number of terminal units, the type of wards, diversity factors and high-demand areas such as operating theatres, ICU, emergency department, delivery suites and treatment rooms. Choosing pumps “by the number of beds” is not enough.

It must be checked whether the system maintains the required negative pressure even when one pump is off, a filter needs replacing or part of the installation is temporarily isolated. The number and type of pumps follow from the design flow and the required redundancy. Often, for example in in.VAC sources, they are selected so that a single pump delivers the full design flow and the others form the reserve.

Where is the air from vacuum pumps discharged?

Air drawn from the vacuum installation may contain biological contaminants, aerosols, moisture and residues from medical procedures. That is why bacterial filters are fitted upstream of the pumps and the exhaust air is discharged safely outside the building.

The exhaust should not be directed into technical rooms, garages, ducts, shafts or attics. Nor should the outlet be located near ventilation air intakes, windows, doors, entrances, terraces, helipads, walkways or places where people stay, where contaminated air could re-enter the building.

The source room needs effective ventilation, because the pumps generate heat, noise and vibration. Too high a temperature shortens pump life and increases the risk of failure.

How should bacterial filters be handled?

Bacterial filters are among the most important elements of a vacuum source and require constant supervision. Their condition is checked, and they are replaced in line with the manufacturer’s instructions, the Operational Management Documentation and the risk assessment.

Used filter cartridges, filter bowls and condensate have been in contact with the contaminated side. Until assessed, they are treated as potentially contaminated material that may need to be handled as medical or hazardous waste. If contamination gets into the pipelines, replacing the filters is not enough, as we explain in the article contaminated medical vacuum pipeline.

Is a vacuum source a medical device?

Yes. According to GMWP, a medical vacuum source must have CE marking and conformity documentation as a medical device or as an element of an assessed system, intended for medical use. Technical performance of the pump alone does not replace this. A vacuum installation is usually classified as a class IIa medical device under Regulation (EU) 2017/745 (MDR), and the final class is determined by the manufacturer. Any intervention in the pipeline is a modification of the device.

One example is the in.VAC source made by INMED S.A., a class IIa medical device with CE marking. Medpipe is its distributor and authorised service provider, and we have gathered the technical data and certificate on the page medical gas supply systems.

Frequently asked questions

What is medical vacuum used for?

Medical vacuum is used to suction fluids, secretions and gases from the treatment field, at intensive care stations, in operating theatres and in other clinical areas.

How many sources must a medical vacuum installation have?

According to GMWP, at least three: primary, secondary and reserve. At least one source or reserve section should be located in another room or another fire compartment.

Does a vacuum tank replace the pump reserve?

No. The buffer tank stabilises the system and smooths short fluctuations in demand, but it does not replace pump capacity or the reserve source.

Where is the exhaust from vacuum pumps discharged?

Outside the building, away from ventilation air intakes, windows, doors, entrances and places where people stay. Bacterial filters are fitted upstream of the pumps.

Can anaesthetic gas scavenging be connected to medical vacuum?

No. The anaesthetic gas scavenging system (AGSS) is a separate system covered by PN-EN ISO 7396-2, with its own flow-generating equipment and terminal units to PN-EN ISO 9170-2. It is not connected to medical vacuum.

How often are the bacterial filters of a vacuum source replaced?

In line with the manufacturer’s instructions, the Operational Management Documentation and the risk assessment. Used cartridges are treated as potentially contaminated material.

Basis

  • Medical Gases in Practice, Booklet 1, 2026 edition, INMED S.A.: chapter 7.4 (example configuration of a VAC source), 7.13 (anaesthetic gas scavenging) and 8.2 (medical gas sources as medical devices)
  • PN-EN ISO 7396-1:2016-07: pipeline systems for compressed medical gases and vacuum
  • PN-EN ISO 9170-1:2020-12: terminal units for compressed medical gases and vacuum
  • PN-EN ISO 7396-2:2011: anaesthetic gas scavenging disposal systems
  • PN-EN ISO 9170-2:2010: terminal units for anaesthetic gas scavenging systems
  • Regulation (EU) 2017/745 on medical devices (MDR)

If you are planning a modernisation or verifying an existing arrangement of supply sources, order the GMWP guide and check your installation against the criteria described in it.

See also: medical compressed air supply source, what the pipeline system certificate does not cover.

Prepared by: technical team of Małgorzata Dopierała, Damian Czyczyro and Przemysław Kostera — Medpipe Sp. z o.o., design, audits and inspections of medical gas installations. The article draws on the guide “Medical gases in practice. Volume 1” (GMWP), INMED S.A., chapters 7.4, 7.13 and 8.2.